Error 2:
The meter has detected an irregularity with the sample.
[1] The test strip may be partially filled; you should check the
window of the test strip to confirm that it is full.
[2] The sample may not be blood or control solution.
Error 3:
The meter has detected that the test strip is in poor condition.
[1] The test strip may have been improperly stored (e.g. hot,
humid conditions) or may be expired.
[2] The test strip may have been mishandled by vigorous
bending or shaking.
[3] Parts of the test strip may have become covered in grease,
oil, or lotion.
[4] The test strip port connector may be dirty.
Error 4:
Unusual test strip problems have occurred that may be
related to extreme conditions.
[1] Retest where it is closer to a room temperature of 21 °C to
24 °C (70 °F to 75 °F).